Can make you unwell for a while. Do not eat it.

MEDLINE Medical Procedure Kits labeled as: 1) CAROTID NEURO PACK, Medline Kit Number/SKU DYNJ44926G; 2) VP SHUNT CDS, Medline Kit Number/SKU CDS983800J; 3) VP SHUNT CDS, Medline Kit Number/SKU CDS983800K.

Check your package for

Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 24EBH848; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 24DBK995; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 23JBP023; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 23HBR067; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 23FBM330; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 23EBK061; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 22JBG042; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 22GBH864; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 22DBT571; Medline Kit Number/SKU CDS983800J: UDI/DI 10195327684662 (each), 40195327684663 (case), Lot Number 25DMJ438; Medline Kit Number/SKU CDS983800J: UDI/DI 10195327684662 (each), 40195327684663 (case), Lot Number 25CMH178; Medline Kit Number/SKU CDS983800J: UDI/DI 10195327684662 (each), 40195327684663 (case), Lot Number 24IME720; Medline Kit Number/SKU CDS983800J: UDI/DI 10195327684662 (each), 40195327684663 (case), Lot Number 24HMG912; Medline Kit Number/SKU CDS983800K: UDI/DI 10198459518973 (each), 40198459518974 (case), Lot Number 26CMC503; Medline Kit Number/SKU CDS983800K: UDI/DI 10198459518973 (each), 40198459518974 (case), Lot Number 26AMB047; Medline Kit Number/SKU CDS983800K: UDI/DI 10198459518973 (each), 40198459518974 (case), Lot Number 25JME109.
Why
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Distributed
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Company
Medline Industries, LP
Announced
Recall began
Read the official notice

Every recall — food, medication, or medical device — is shown word-for-word from the official FDA or USDA notice, with a link straight to it. Nothing here is summarised, reworded, or filled in. This site is not affiliated with any government agency.