Español
TemporaryFDAOngoing

Can make you unwell for a while. Do not eat it.

MEDLINE Medical Procedure Kits labeled as: 1) KIT UROLOGY URETHROPLASTY, Medline Kit Number/SKU DYKMBNDL117C; 2) URETHROPLASTY, Medline Kit Number/SKU DYNJ908978; 3) URETHROPLASTY, Medline Kit Number/SKU DYNJ908978F.

Check your package for

Medline Kit Number/SKU DYKMBNDL117C: UDI/DI 10195327371104 (each), 40195327371105 (case), Lot Number 23DLA834; Medline Kit Number/SKU DYNJ908978: UDI/DI 10195327231132 (each), 40195327231133 (case), Lot Number 23BMC463; Medline Kit Number/SKU DYNJ908978: UDI/DI 10195327231132 (each), 40195327231133 (case), Lot Number 23AMA685; Medline Kit Number/SKU DYNJ908978: UDI/DI 10195327231132 (each), 40195327231133 (case), Lot Number 23AMB905; Medline Kit Number/SKU DYNJ908978: UDI/DI 10195327231132 (each), 40195327231133 (case), Lot Number 22LMG466; Medline Kit Number/SKU DYNJ908978: UDI/DI 10195327231132 (each), 40195327231133 (case), Lot Number 22LMB983; Medline Kit Number/SKU DYNJ908978F: UDI/DI 10198459228803 (each), 40198459228804 (case), Lot Number 25JBD225; Medline Kit Number/SKU DYNJ908978F: UDI/DI 10198459228803 (each), 40198459228804 (case), Lot Number 25HBW236.
Why
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Distributed
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Company
Medline Industries, LP
Announced
Recall began
Read the official notice

Every recall — food, medication, or medical device — is shown word-for-word from the official FDA or USDA notice, with a link straight to it. Nothing here is summarised, reworded, or filled in. This site is not affiliated with any government agency.

Local Recall