Español
TemporaryFDAOngoing

Can make you unwell for a while. Do not eat it.

MEDLINE Urethral Catheter, Red Rubber Latex, Medline Product Number/SKU (REF), French sizes: 1. DYND13508 - 08 Fr; 2. DYND13510 - 10 Fr; 3. DYND13512 - 12 Fr; 4. DYND13514 - 14 Fr; 5. DYND13515 - 15 Fr; 6. DYND13516 - 16 Fr; 7. DYND13518 - 18 Fr; 8. DYND13520 - 20 Fr; 9. DYND13522 - 22 Fr; 10. DYND13614 - 14 Fr; 11. DYND13616 - 16 Fr.

Check your package for

1. DYND13508, UDI/DI 10080196031139 (each), 20080196031136 (case); 2. DYND13510, UDI/DI 10080196726462 (each), 40080196726463 (case); 3. DYND13512, UDI/DI 10080196031177 (each), 40080196031178 (case); 4. DYND13514, UDI/DI 10080196726219 (each), 40080196726210 (case); 5. DYND13515, UDI/DI 10080196031214 (each), 40080196031215 (case); 6. DYND13516, UDI/DI 10080196031221 (each), 40080196031222 (case); 7. DYND13518, UDI/DI 10080196031245 (each), 40080196031246 (case); 8. DYND13520, UDI/DI 10080196031252 (each), 40080196031253 (case); 9. DYND13522, UDI/DI 10080196061259 (each), 40080196061250 (case); 10. DYND13614, UDI/DI 10080196031276 (each), 30080196031270 (case); 11. DYND13616, UDI/DI 10888277720879 (each), 20888277720876 (case); ALL LOTS
Why
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Distributed
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Company
Medline Industries, LP
Announced
Recall began
Read the official notice

Every recall — food, medication, or medical device — is shown word-for-word from the official FDA or USDA notice, with a link straight to it. Nothing here is summarised, reworded, or filled in. This site is not affiliated with any government agency.

Local Recall