Español
TemporaryFDAOngoing

Can make you unwell for a while. Do not eat it.

Ritmed Neurological Sponges, 0.75x0.75in, Sterile, REF: 85995C

Check your package for

UDI-DI: 10686864044465, Lot: 17936
Why
Neurological sponges used to moisten or absorb biological fluid during neurological surgery and may also be used to protect the brain during incision, may not consistently meet endotoxin specification for devices that contact the cerebrospinal fluid. Elevated endotoxin levels may lead to neurovascular adverse events, febrile response, and/or local transient inflammation, hypotension, or nausea.
Distributed
US: MT, NC, VA, CA, TX, OH, TN, IL, RI
Company
AMD Medicom Inc.
Announced
Recall began
Read the official notice

Every recall — food, medication, or medical device — is shown word-for-word from the official FDA or USDA notice, with a link straight to it. Nothing here is summarised, reworded, or filled in. This site is not affiliated with any government agency.

Local Recall