Can make you unwell for a while. Do not eat it.

Ritmed Neurological Sponges, Description/REF: 0.5x0.5in, Non-Sterile/55992C; 0.5x1.0in, Non-Sterile/55993C; 0.5x3.0in, Non-Sterile/55994C; 0.75x0.75in, Non-Sterile/55995C; 1.0x1.0in, Non-Sterile/55996C; 1.0x3.0in, Non-Sterile/55997C; 0.5x1.5in, Non-Sterile/55998C

Check your package for

UDI-DI: 20686864043205. Distributed After May 1st, 2022, including the following lots: 19316, 16409, 15463-7, 15463-6, 15463-5, 15463-4, 15463-3, 15463-2, 15463-1, 15463, 15192-1, 15192, 14813, 14039, 13673, 13289, 12641, 12206, 11851, 11020-1, 11020, 10586, 10392, 9243, 8160, 7784, 7748, 7044
Why
Neurological sponges used to moisten or absorb biological fluid during neurological surgery and may also be used to protect the brain during incision, may not consistently meet endotoxin specification for devices that contact the cerebrospinal fluid. Elevated endotoxin levels may lead to neurovascular adverse events, febrile response, and/or local transient inflammation, hypotension, or nausea.
Distributed
US: MT, NC, VA, CA, TX, OH, TN, IL, RI
Company
AMD Medicom Inc.
Announced
Recall began