TemporaryFDAOngoing
Can make you unwell for a while. Do not eat it.
TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816
Check your package for
UDI-DI: 03596010514394. Lot/Batch: 25BM03616
- Why
- Humeral nail due to a manufacturing error is missing distal holes and if implanted may lead to revision surgery but, if the missing holes are identified before implantation, an equivalent backup can be used, or a bent nail of the same size, which would require additional surgical steps to address soft tissue/bone preparation, which could lead to a surgical delay of greater than 30 minutes.
- Distributed
- International distribution to the countries of BE, DE, IT.
- Company
- Smith & Nephew, Inc.
- Announced
- Recall began