Bicarby Dialysate; Model number: RFP-400-G;

Check your package for

Model number: RFP-400-G; UDI-DI: 00840861102839; Lot codes: 25SG06026, 25SG06027;
Why
There is an error in the labeling of the GTIN on the product. When scanned; however, the case labeling, lot number, and printed labeling on the product correctly identifies the product code and the solution contained in the bag. In the event the incorrect GTIN labeling is used as the singular source for product identification, unrecognized use of a potassium solution with a concentration differing from the prescription may result.
Distributed
Domestic: TX.
Company
Fresenius Medical Care Holdings, Inc.
Announced