Veritas Reverse Total Shoulder Replacement, Model/Catalog Number: 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1, size 1 -2, 40mm Polymer Liner

Check the device or its packaging for

All Lot codes/ UDIs:Product Code: 3500-140C3 UDI/GTIN:00840097513492; Product Code: 3500-140U3 UDI/GTIN:00840097513461; Product Code: 3500-140C1 UDI/GTIN:00840097513430; Product Code: 3500-140U1 UDI/GTIN:00840097513409
Why
(Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or revision surgery.
Distributed
US Nationwide distribution in the states of Texas, Massachusetts, Oklahoma, Florida, Colorado, North Carolina.
Company
restor3d Inc.