Español
TemporaryFDAOngoing

Can cause harm that is usually temporary.

Mobile column TruSystem 7500 U, Product Code 1730720

Check the device or its packaging for

UDI-DI: 00887761974388; Serial number: 110183858
Why
The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.
Distributed
US Nationwide distribution in the states of IL, MO, TX.
Company
Baxter Healthcare Corporation
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-0333-2027 · retrieved 3 November 2026

Read the official notice

Every recall is shown word for word from the FDA and USDA, with a link to the source.

Local Recall