SeriousFDAOngoing
Can cause serious harm or death.
Impella Optical Controller, Packaged. Primary user control interface for the Impella Catheter, a catheter-based heart pump.
Check the device or its packaging for
Product Code: 0042-0010-US, 0042-0010-UK, 0042-0010-EU, 0042-0010-IN; UDI: 00813502010985, 00813502010978, 00813502010961, 00813502011425; Serial No. IC6962, IC7004, IC7036, IC7170, IC7213, IC7214, IC7239, IC7241, IC7677, IC7787, IC7880, IC8105, IC8106, IC8295, IC8308, IC8315, IC8316, IC8335, IC8541, IC8543, IC11456, IC7937, IC9832, IC9833, IC5409, IC7798, IC7800, IC7946, IC8070, IC8337, IC8441, IC7825, IC8126, IC8630, IC8640, IC9804, IC11451, IC11497, IC8388, IC8751, IC9501, IC9989, IC10102, IC10103, IC10318, IC10589, IC10463, IC10979, IC11299, IC8164, IC9348, IC9500, IC9581, IC9648, IC9777, IC9793, IC9797, IC9915, IC9919, IC9922, IC10278, IC10302, IC10983, IC10984, IC10985, IC11174, IC11410, IC13231, IC13263, IC13264, IC13265, IC13681, IC13682, IC10040, IC10043, IC8401, IC8748, IC8980, IC8609, IC9178, IC9177, IC9324, IC9328, IC9557, IC9558, IC9744, IC9745, IC9746, IC9908, IC9921, IC9969, IC9970, IC10063, IC10064, IC10149, IC10150, IC10238, IC10239, IC10341, IC10366, IC10367, IC12407, IC12408, IC12409, IC13057, IC13060, IC13061, IC13064, IC8333, IC8494, IC12326, IC4952, IC5041, IC5158, IC5281, IC5283, IC5284, IC5403, IC5404, IC5413, IC5527, IC5588, IC5635, IC5636, IC5641, IC5643, IC5666, IC5667, IC5804, IC5805, IC5918, IC5921, IC6013, IC6343, IC6421, IC6422, IC6461, IC6463, IC6465, IC6777, IC6785, IC6944, IC6945, IC6946, IC7161, IC7233, IC7232, IC7235, IC7626, IC7738, IC7739, IC7792, IC7793, IC7826, IC7956, IC8012, IC8258, IC8423, IC8424, IC8499, IC8511, IC8723, IC8596, IC8665, IC8765, IC7827, IC7828, IC8404, IC11746, IC12344, IC12327, IC8426, IC13232, IC13698, IC7861, IC7862, IC8121, IC8222, IC8225.
- Why
- Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. 2. Potential for the rotary push button (RPB) sticking and unreliable actuation impacting operator ability to select and confirm menu entries. This can result in loss of control of the console. 3. Potential AIC instability that could result in AIC start-up failure, inability to drive the pump, and the need for console exchange. 4. Software defect related to calculator process crash, leading to Negative Central Venous Pressure (CVP) values present in Pulmonary Artery Pressure (PAP) calculations.
- Distributed
- Worldwide Distribution: US (nationwide) and OUS (foreign) to countries: Belgium, Brazil, Canada, Colombia, Croatia, Czech Republic, Finland, France, Germany, Hong Kong, India, Ireland, Italy, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Pakistan, Panama, Poland, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom
- Company
- Abiomed, Inc.
- Announced
- This is the date the recall was announced. A company may have started it weeks earlier.
- Recall began
Source: FDA record Z-0209-2027 · retrieved 28 October 2026
Read the official notice