Español
TemporaryFDAOngoing

Can cause harm that is usually temporary.

TS Misc. Instruments Tray, Triathlon. Catalog Number: 6543-6-519.

Check the device or its packaging for

Catalog Number: 6543-6-519. GTIN: 07613327555127. All lot numbers used prior to the correction date of May 8, 2026 are impacted.
Why
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
Distributed
These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.
Company
Howmedica Osteonics Corp.
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-0167-2027 · retrieved 15 October 2026

Read the official notice

Every recall is shown word for word from the FDA and USDA, with a link to the source.