Can cause harm that is usually temporary.

Triathlon Hinge Femoral Prep Tray. Catalog Number: 6543-6-514.

Check the device or its packaging for

Catalog Number: 6543-6-514. GTIN: 07613327562859. All lot numbers used prior to the correction date of May 8, 2026 are impacted.
Why
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
Distributed
These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.