TS Universal Baseplate Finishing Tray, Triathlon. Catalog Number: 6543-6-506.
Check the device or its packaging for
Catalog Number: 6543-6-506. GTIN: 07613327548082. All lot numbers used prior to the correction date of May 8, 2026 are impacted.
- Why
- Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
- Distributed
- These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.
- Company
- Howmedica Osteonics Corp.