LowFDAOngoing
Unlikely to cause harm, but it breaks the rules.
LeadCare II Blood Lead Test Kit. Model/Catalog Number: 70-6762.
Check the device or its packaging for
Model/Catalog Number: 70-6762. UD-DI: 00850355006000. Lot Number. 2530M-01.
- Why
- In kit sub-lot 2530M-01 the date of expiration printed on the outer labeling may indicate the wrong expiration date of July 22, 2026 while the correct expiration date is July 13, 2026. If used the system will display an error message and prevent the operator from running the test. Resulting in delayed results.
- Distributed
- US Nationwide distribution in the states of CA, IN, PA, TX, NE, IL, MA, MO, KY.
- Company
- Magellan Diagnostics, Inc.
- Announced
- This is the date the recall was announced. A company may have started it weeks earlier.
- Recall began
Source: FDA record Z-0162-2027 · retrieved 15 October 2026
Read the official notice