TemporaryFDAOngoing
Can cause harm that is usually temporary.
RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115
Check the device or its packaging for
UDI: 00724995245658 Lot numbers with expiration dates: 687046 10/5/2027 688017 10/19/2027
- Why
- If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.
- Distributed
- Worldwide - US Nationwide distribution.
- Company
- Steris Corporation
- Announced
- This is the date the recall was announced. A company may have started it weeks earlier.
- Recall began
Source: FDA record Z-0154-2027 · retrieved 14 October 2026
Read the official notice