Español
TemporaryFDAOngoing

Can cause harm that is usually temporary.

RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115

Check the device or its packaging for

UDI: 00724995245658 Lot numbers with expiration dates: 687046 10/5/2027 688017 10/19/2027
Why
If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.
Distributed
Worldwide - US Nationwide distribution.
Company
Steris Corporation
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-0154-2027 · retrieved 14 October 2026

Read the official notice

Every recall is shown word for word from the FDA and USDA, with a link to the source.

Local Recall