Can cause serious harm or death.

V+ Pro Emergency, Model Number: PRT-01201-000, Package PN:PRT-01173-000,PRT-01222-000. used for ventilatory support.

Check the device or its packaging for

All Serial Numbers: UDI: 00855573007945
Why
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Distributed
Distribution US nationwide including Puerto Rico.
Company
Ventec Life Systems, Inc.