Español
TemporaryFDAOngoing

Can cause harm that is usually temporary.

Rollator, Bariatric, Basic Steel, Burgundy, mechanical walker with four legs and a metal frame intended for medical purposes to provide moderate weight support while walking, Product Number MDS86800XWS

Check the device or its packaging for

Lot Numbers: 05425090001 05425120001 05426010001 05426020001 05426030001 05426040001 05426050001
Why
The front fork stem bolt/wheel assembly may disconnect from the rollator frame tubing. If the front fork stem bolt/wheel assembly disconnects from the rollator frame tubing, the front wheels may become loose or detach from the rollator. If the front wheels become loose or detach during use, there is risk of severe adverse events. The immediate consequences could include falls resulting in head trauma, bone fractures, or other critical injuries.
Distributed
Worldwide distribution - US Nationwide and the country of Canada.
Company
Medline Industries, LP
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-0086-2027 · retrieved 5 October 2026

Read the official notice

Every recall is shown word for word from the FDA and USDA, with a link to the source.