TemporaryFDAOngoing
Can cause harm that is usually temporary.
Brand Name: MPR Product Name:SINGLE USE TOURNIQUET CUFF 18" 1 LINE (10) Model/Catalog Number: DT-7018-01Q Software Version: Not Applicable Product Description: Sterile Disposable Tourniquet Cuff is used to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery. It is non-invasive device. Tourniquet Cuffs are intended for use in surgical procedures that requires the interruption of blood flow and clear visualization of the operative field. Component: No
Check the device or its packaging for
Lot Code: Model Number: DT-7018-01Q UDI-DI: UDI was not applicable at time of manufacture. All Lots produced from 01SEP2016 to 01MAR2023: S1101STY, Q0301STY, P1015STY, N0330STY, N0530STY
- Why
- Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
- Distributed
- US Nationwide distribution in the states of MO, NV, IN, MN, GA, WI, PA, TX, CT, LA, NJ, FL, AL, NY, SC, NE, NC, OH, AZ, CO, MA, IL, KY, CA, VA, ME, KS, MD, TN, MS, AR, WA, NM, MI, WY.
- Company
- ASP Global, LLC.
- Announced
- This is the date the recall was announced. A company may have started it weeks earlier.
- Recall began
Source: FDA record Z-0052-2027 · retrieved 1 October 2026