Español
TemporaryFDAOngoing

Can cause harm that is usually temporary.

Brand Name: ASP Global Product Name:DISPOSABLE TOURNIQUET CUFF, 30" X 4" 1-Line, White Model/Catalog Number: ASP-564010 Software Version: Not Applicable Product Description: Sterile Disposable Tourniquet Cuff is used to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery. It is non-invasive device. Tourniquet Cuffs are intended for use in surgical procedures that requires the interruption of blood flow and clear visualization of the operative field. Component: No

Check the device or its packaging for

Lot Code: Model Number: ASP-564010 UDI-DI: (01) 00840216106758 All Lots produced from 01SEP2016 to 01MAR2023: P1015STY, P1215STY, O1020STY, N0330STY, N0530STY, M0130STY, M0430STY, M0730STY, M0930STY, M1015STY, M1101STY, M1115STY, L0228STY, L0310STY, L0325STY
Why
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Distributed
US Nationwide distribution in the states of MO, NV, IN, MN, GA, WI, PA, TX, CT, LA, NJ, FL, AL, NY, SC, NE, NC, OH, AZ, CO, MA, IL, KY, CA, VA, ME, KS, MD, TN, MS, AR, WA, NM, MI, WY.
Company
ASP Global, LLC.
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-0043-2027 · retrieved 1 October 2026

Read the official notice

Every recall is shown word for word from the FDA and USDA, with a link to the source.