Brand Name: Expressaire Product Name: DISPOSABLE TOURNIQUET CUFF, 34" X 4" 1-LINE (CS 10) BLUE Model/Catalog Number: 564012 Software Version: Not Applicable Product Description: Sterile Disposable Tourniquet Cuff is used to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery. It is non-invasive device. Tourniquet Cuffs are intended for use in surgical procedures that requires the interruption of blood flow and clear visualization of the operative field. Component: No

Check the device or its packaging for

Lot Code: Model Number: 564012 UDI-DI: (01) 00885914044629 All Lots produced from 01SEP2016 to 01MAR2023: S0922STY, S1101STY, R0601STY, R1101STY, Q0301STY, Q1115STY, P0325STY, P0620STY, P1015STY, O1020STY, N0730STY, M0130STY, M0430STY, M0730STY, M0930STY, M1230STY
Why
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Distributed
US Nationwide distribution in the states of MO, NV, IN, MN, GA, WI, PA, TX, CT, LA, NJ, FL, AL, NY, SC, NE, NC, OH, AZ, CO, MA, IL, KY, CA, VA, ME, KS, MD, TN, MS, AR, WA, NM, MI, WY.