TemporalFDAEn curso

Puede causar daños que por lo general son pasajeros.

SCORPIO + MOBILE BEARING PCS 78-5-0910 SYSTEM ONE BIPOLAR 40MM 6074-4022 SYSTEM ONE BIPOLAR 43MM 6074-4322 SYSTEM ONE BIPOLAR 44MM 6074-4422 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

La agencia publicó este retiro solo en inglés. El nombre del producto y los códigos del paquete se muestran tal como los escribió la agencia, para que coincidan exactamente con su paquete.

Revise el dispositivo o su empaque

78-5-0910 UDI-DI 07613327366006 Lot 30582401 6074-4022 UDI-DI 07613327031775 Lots 2M5NTY LH25R5 M36N83 6074-4322 UDI-DI 07613327031805 Lot W40E3T 6074-4422 UDI-DI 07613327031812 Lot R84K20
Motivo
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.La agencia escribió este motivo solo en inglés.La agencia escribió este motivo solo en inglés.
Distribuido
Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
Empresa
Howmedica Osteonics Corp.
Anunciado
Esta es la fecha en que se anunció el retiro. La empresa pudo haberlo iniciado semanas antes.
El retiro comenzó

Fuente: registro Z-3205-2026 de FDA · consultado el 16 de septiembre de 2026