BEARING AND AXLE BUSHING ASSY STD ELBOW C-M113-1-301 KRH REV TIB INS 16mm THK .640 DIA STEM C-M113-5-600 SELF-PRESSURIZING GLENOID SIZE 9 C-M063-5-800 Orthopaedic implant components manufactured from UHMWPE (polyethylene)
La agencia publicó este retiro solo en inglés. El nombre del producto y los códigos del paquete se muestran tal como los escribió la agencia, para que coincidan exactamente con su paquete.
Revise el dispositivo o su empaque
C-M113-1-301 UDI-DI 07613327590074 Lot MH621A C-M113-5-600 No UDI Lot 194593 C-M063-5-800 No UDI Lot RX06358A
- Motivo
- Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.La agencia escribió este motivo solo en inglés.La agencia escribió este motivo solo en inglés.
- Distribuido
- Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.