Puede causar daños graves o la muerte.

BD Needle Kit for Powered Driver 15mm x 15Ga, REF: D015151NK, BD Needle Kit for Powered Driver 25mm x 15Ga, REF: D015251NK, BD Needle Kit for Powered Driver 35mm x 15Ga, REF: D015351NK, BD Needle Kit for Powered Driver 45mm x 15Ga, REF: D015451NK, BD Needle Kit for Powered Driver 55mm x 15Ga, REF: D015551NK

La agencia publicó este retiro solo en inglés. El nombre del producto y los códigos del paquete se muestran tal como los escribió la agencia, para que coincidan exactamente con su paquete.

Revise el dispositivo o su empaque

REF/UDI-DI/Lot(Expiration): D015151NK/00801741163586/BSLC9275(2026-08-03), BSLC9563(2026-08-13), BSLC9567(2026-08-25), BSLC10103(2026-09-27), BSLC10516(2026-11-01), BSLC10526(2026-11-16), BSLC10872(2026-12-09), BSLC11216(2028-07-02), BSLC11456(2028-07-23); D015251NK/00801741163678/BSLC9467(2026-08-10), BSLC9575(2026-09-20), BSLC10105(2026-10-04), BSLC10522(2026-11-07), BSLC10868(2026-12-02), BSLC11212(2028-06-18), BSLC11454(2028-07-17), BSLC11463(2028-08-07), BSLC11884(2028-09-15), BSLC11890(2028-09-18), BSLC12082(2028-10-08), BSLC12115(2028-10-20), BSLC12123(2028-10-22); D015351NK/00801741163593/BSLC11458(2028-07-24); 00801741163678/BSLC9277(2026-08-04), BSLC9485(2026-08-12), BSLC9569(2026-08-26), BSLC10099(2026-09-07), BSLC10514(2026-10-16), BSLC10518(2026-10-21), BSLC10524(2026-11-08), BSLC10870(2026-12-03), BSLC10874(2028-06-12), BSLC11218(2028-07-07); D015451NK/00801741163609/BSLC9279(2026-08-06), BSLC9565(2026-08-17), BSLC9571(2026-09-07), BSLC10101(2026-09-17), BSLC10456(2026-10-14), BSLC10520(2026-10-22), BSLC10866(2026-11-21), BSLC11214(2028-06-23), BSLC11450(2028-07-09), BSLC11460(2028-07-31), BSLC11873(2028-08-13), BSLC11888(2028-09-25), BSLC12007(2028-10-02), BSLC12105(2028-10-15); D015551NK/00801741163616/BSLC9465(2026-08-07), BSLC9573(2026-09-26), BSLC10107(2026-10-02), BSLC10528(2026-11-20), BSLC11452(2028-07-10)
Motivo
After placement of intraosseous (IO) needle sets, used when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases, users may experience difficulty/inability to remove obturator, due to it locking to the needle hub, due to out-of-tolerance dimensions and IFU unclarity, which may result in new needle placement, which may result in delayed therapyLa agencia escribió este motivo solo en inglés.La agencia escribió este motivo solo en inglés.
Distribuido
US: MA, FL, IA, SC, IL, NY, UT, NJ, ID, IN, PA, CA, TX, OH, MO, VA, MD, NC, AL, MN, CO, KY, LA, NV, OK, MS, AZ, MI, GA, AR, KS, TN. OUS: CA, SG
Empresa
Bard Access Systems, Inc.
Anunciado
Esta es la fecha en que se anunció el retiro. La empresa pudo haberlo iniciado semanas antes.
El retiro comenzó

Fuente: registro Z-2872-2026 de FDA · consultado el 14 de agosto de 2026

Leer el aviso oficial

Cada retiro — de alimentos, medicamentos o dispositivos médicos — se muestra palabra por palabra del aviso oficial de la FDA o del USDA, con un enlace directo. Aquí nada se resume, se reescribe ni se rellena. Este sitio no está afiliado a ninguna agencia del gobierno.

Local Recall