TemporalFDAEn curso
Puede enfermarle por un tiempo. No lo coma.
FARADRIVE STEERABLE SHEATH CLEAR OUS, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PFCE21M402, IFU Reference Number 51659727-01
La agencia publicó este retiro solo en inglés. El nombre del producto y los códigos del paquete se muestran tal como los escribió la agencia, para que coincidan exactamente con su paquete.
Revise su paquete
UDI-DI: 00191506042516; Lots: CL12917, CL13004, CL13017, CL13046, CL13060, CL13080, CL13103, CL13122, CL13123, CL13142, CL13160, CL13196, CL13212, CL13219, CL13231, CL13260, CL13291, CL13340, CL13375, CL13401, CL13422, CL13468, CL13519, CL13573, CL13578, CL13614, CL13646, CL13670, CL13671, CL13708, CL13719, CL13722, CL13732, CL13760, CL13801, CL13802, CL13820, CL13833, CL13856, CL13910, CL13911, CL13925, CL13940, CL13971, CL13972, CL13988, CL14025, CL14039, CL14050, CL14051, CL14076, CL14077, CL14119, CL14120, CL14135, CL14137, CL14171, CL14187, CL14226, CL14249, CL14286, CL14287, CL14345, CL14346, CL14375, CL14376, CL14396, CL14397, CL14439, CL14488, CL14500, CL14501, CL14518, CL14519, CL14581, CL14582, CL14633, CL14635, CL14670, CL14697, CL14711, CL14749, CL14750, CL14769, CL14770, CL14813, CL14814, CL14854, CL14855, CL14870, CL14871, CL14928, CL14929, CL14933, CL14934, CL14935, CL14968, CL14969, CL15005, CL15009, CL15067, CL15107, CL15108, CL15109, CL15128, CL15129, CL15191, CL15192, CL15193, CL15239, CL15240, CL15272, CL15318, CL15358, CL15397, CL15427, CL15474, CL15490, CL15539, CL15568, CL15569, CL15597, CL15598, CL15637, CL15638, CL15640, CL15754, CL15755, CL15775, CL15776, CL15808
- Motivo
- Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
- Distribuido
- Global Distribution. US Nationwide.
- Empresa
- Boston Scientific Corporation
- Anunciado
- El retiro comenzó