TemporalFDAEn curso
Puede causar daños que por lo general son pasajeros.
Power Over Fiber Base Module
La agencia publicó este retiro solo en inglés. El nombre del producto y los códigos del paquete se muestran tal como los escribió la agencia, para que coincidan exactamente con su paquete.
Revise el dispositivo o su empaque
Power Over Fiber Base Module Identification P/N 633232 P/N 622365 P/N 633383 P/N 633910 P/N 635731
- Motivo
- Device not certified - no laser product report was filed with CDRH prior to distribution. Product label incorrectly states device is "Class IV", although Class 1 is appropriate for device. Improperly labeled - omitted labels required under 21 CFR 1040.10 / IEC 60825-1 via LN#56. Missing certification and proper manufacturer's ID label. Was not accompanied by compliant User Information. Formalized quality control procedures and record keeping of QC testing was not performed to IEC 60825-1 nor 21 CFR 1040.10 standards. Sales information did not convey required class designation and warning statementLa agencia escribió este motivo solo en inglés.La agencia escribió este motivo solo en inglés.
- Distribuido
- U.S. Nationwide Distribution and International Distribution
- Empresa
- Terex Utilities
- Anunciado
- Esta es la fecha en que se anunció el retiro. La empresa pudo haberlo iniciado semanas antes.
- El retiro comenzó
Fuente: registro Z-2745-2026 de FDA · consultado el 30 de julio de 2026