Omnipod DASH Pod. Includes the below Model/REF Numbers: 1. POD-BLE-I1-525 (5-Pack), POD-BLE-I1-520 (Single Pod). 2. POD-BLE-I1-529 (10-Pack), POD-BLE-I1-520 (Single Pod). 3. BLE-I1-529 (10-Pack), BLE-I1-520 (Single Pod). 4. BLE-C1-529 (10-Pack), BLE-C1-520 (Single Pod).

La agencia publicó este retiro solo en inglés. El nombre del producto y los códigos del paquete se muestran tal como los escribió la agencia, para que coincidan exactamente con su paquete.

Revise su paquete

1. Model/REF Number: POD-BLE-I1-525 (5-Pack), POD-BLE-I1-520 (Single Pod). Pack UDI-DI: 20385082000051. Single Pod UDI-DI: 10385082000023. Lot Numbers: PD1U11052421, PD1U11012421, PD1U11022421, PD1U02152521, PD1U02192521, PD1U04242522, PD1U10282522, PD1U10292521, PD1U10302521, PD1U10312521, PD1U11032521, PD1U11042521, PD1U01082621. 2. Model/REF Number: POD-BLE-I1-529 (10-Pack), POD-BLE-I1-520 (Single Pod). Pack UDI-DI: 20385082000020. Single Pod UDI-DI: 10385082000023. Lot Numbers: PD1U11192421, PD1U09102421, PD1U09092421, PD1U09062421, PD1U10242421, PD1U10252421, PD1U10282421, PD1U10292421, PD1U10312421, PD1U11042421. 3. Model/REF Number: BLE-I1-529 (10-Pack), BLE-I1-520 (Single Pod). Pack UDI-DI: 20385082000020. Single Pod UDI-DI: 10385082000023. Lot Numbers: PD1U11152421, PD1U11182421, PD1U11202421, PD1U09042421, PD1U11132421, PD1U11122421, PD1U11142421, PD1U11152411, PD1U11162421, PD1U04292521, PD1U09032421, PD1U11112421, PD1U04262521. 4. Model/REF Number: BLE-C1-529 (10-Pack), BLE-C1-520 (Single Pod). Pack UDI-DI: 20385082000112. Single Pod UDI-DI: 10385082000115. Lot Numbers: PD1U03042521, PD1U09052421, PD1U11072421, PD1U11082421, PD1U03062521, PD1U03032521, PD1U03072521, PD1U02282525, PD1U02282526, PD1U04252521, PD1U11062521, PD1U11072521, PD1U11102521, PD1U11112511, PD1U11112522, PD1U11122521.
Motivo
External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.
Distribuido
Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, Switzerland, Germany, Denmark, Finland, France, UK, Greece, Israel, Italy, Kuwait, Leichtenstein, Netherlands, Norway, Saudi Arabia, Sweden and Turkey.
Empresa
Insulet Corporation
Anunciado
El retiro comenzó
Leer el aviso oficial

Cada retiro — de alimentos, medicamentos o dispositivos médicos — se muestra palabra por palabra del aviso oficial de la FDA o del USDA, con un enlace directo. Aquí nada se resume, se reescribe ni se rellena. Este sitio no está afiliado a ninguna agencia del gobierno.

Local Recall