GE Healthcare AW Server 3.2 ext.6.5; Radiological Image Processing System
La agencia publicó este retiro solo en inglés. El nombre del producto y los códigos del paquete se muestran tal como los escribió la agencia, para que coincidan exactamente con su paquete.
Revise su paquete
REF# (01)00840682102384(10)AWS03D02E6D5, GTIN Number 00840682102384
- Motivo
- GE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches an interactive application (e.g. Volume Viewer), the application may open the previous patient's exam instead of the intended one. When this issue occurs, there is no system warning or error notification. If the error is not recognized, a clinical user could review, interpret, or report images for the wrong patient which could lead to misdiagnosis, incorrect clinical decisions, resulting in delayed or incorrect treatment.
- Distribuido
- Worldwide distribution - United States Nationwide and the countries of Algeria, Australia, Belarus, Belgium, Benin, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Korea (Republic Of), Lithuania, Morocco, Nepal, Netherlands, New Zealand, North Macedonia, Norway, Pakistan, Poland, Portugal, Romania, Russia, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkiye, United Arab Emirates, United Kingdom, Vietnam.