English
TemporalFDAEn curso

Puede enfermarle por un tiempo. No lo coma.

Brand Name: HeartMate Product Name: HM 11V Li-ion Backup Battery, ABT Model/Catalog Number: 106128 Software Version: N/A Product Description: The HeartMate 11 Volt Lithium Ion Standalone Backup Battery is a component of the HeartMate 3" System Controller and provides power to the System Controller. The HeartMate System Controller is an external device that manages the HeartMate Left Ventricular Assist System (LVAS), including alarms, system monitoring, and communication. It contains an internal 11V lithium-ion backup battery that provides at least 15 minutes of emergency power should the main system power source fail. The backup battery must be properly installed, charges automatically during normal use, and has a 36-month service life from date of manufacture. Component: Yes, HeartMate 3" System Controller

La agencia publicó este retiro solo en inglés. El nombre del producto y los códigos del paquete se muestran tal como los escribió la agencia, para que coincidan exactamente con su paquete.

Revise su paquete

Lot Code: Model Number: 106128 UDI/GTIN 05415067039880 Serial Numbers Distributed in US: See Attachment E for serial numbers distributed.
Motivo
Due to 11 volt Backup Battery failures.
Distribuido
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI and the countries of Canada, Germany, Japan, Netherlands, Poland, Spain.
Empresa
Thoratec LLC
Anunciado
El retiro comenzó
Leer el aviso oficial

Cada retiro — de alimentos, medicamentos o dispositivos médicos — se muestra palabra por palabra del aviso oficial de la FDA o del USDA, con un enlace directo. Aquí nada se resume, se reescribe ni se rellena. Este sitio no está afiliado a ninguna agencia del gobierno.