English
GraveFDAEn curso

Puede causar una enfermedad grave o la muerte. No lo coma.

Description/REF: HEMODIALYSIS KIT: 2-LUMEN 12 FR X 20 CM/AK-15122-F, AHDC KIT: 3L 12 FR X 16 CM/ASK-12123-UOC1, AHDC KIT: 3-L 12 FR X 16 CM/ASK-12123-UPM, AHDC KIT: 3L 12 FR X 20 CM/ASK-15123-UOC1, AHDC KIT: 3-L 12 FR X 20 CM/ASK-15123-UPM, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-P1A, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-XP1A, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-XPCN1A, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-XPN1A, AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-P1A, AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-XPCN1A, AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-XPN1A, AHDC KIT: 3-L 12 FR X 25 CM/CDC-16123-XPCN1A, HEMODIALYSIS KIT 2L: 12 FR X 20CM AGB/AK-25122-CDC, HEMODIALYSIS KIT: 2-LUMEN 14 FR X 20 CM/AK-25142-F, HEMODIALYSIS KIT: 2L 12 FR X 6 IN/ASK-22122-MM1, HEMODIALYSIS KIT: 2L 12 FR X 6 IN/ASK-22122-MM2, AHDC KIT: 2L 14 FR X 15 CM/ASK-22142-UOC1, AHDC AGB KIT: 2-L 14 FR X 15 CM/ASK-22142-UPM, HEMODIALYSIS KIT: 2L 14 FR X 20 CM/ASK-25142-UNC1, AHDC KIT: 2L 14 FR X 20 CM/ASK-25142-UOC1, AHDC AGB KIT: 2L 14 FR X 20 CM/ASK-25142-UPM, HEMODIALYSIS KIT 2L: 14 FR X 25 CM AGB/ASK-26142-UNC1, HEMODIALYSIS KIT: 2LUMEN 12 FR X 16 CM/AU-22122-F, HEMODIALYSIS KIT: 2L 12 FR X 16 CM AGB/CDC-22122-1A, AHDC AGB U-BEND KIT: 2-L 12 FR X 16 CM/CDC-22122-XCN1A-U, AHDC AGB KIT: 2-L 12 FR X 16 CM/CDC-22122-XN1A, AHDC AGB U-BEND KIT: 2-L 12 FR X 16 CM/CDC-22122-XN1A-U, HEMODIALYSIS KIT 2L: 14 FR X 15 CM AGB/CDC-22142-1A, HEMODIALYSIS KIT: 2L 14 FR X 15 CM AGB/CDC-22142-XC1A, AHDC AGB KIT: 2-L 14 FR X 15 CM/CDC-22142-XCN1A, AHDC AGB CURVED KIT: 2-L 14 FR X 15 CM/CDC-22142-XCN1A-C, AHDC AGB KIT: 2-L 14 FR X 15 CM/CDC-22142-XN1A, AHDC AGB CURVED KIT: 2-L 14 FR X 15 CM/CDC-22142-XN1A-C, AHDC AGB U-BEND KIT: 2-L 12 FR X 13 CM/CDC-23122-XCN1A-U, AHDC AGB U-BEND KIT: 2-L 12 FR X 13 CM/CDC-23122-XN1A-U, HEMODIALYSIS KIT: 2L 12 FR X 13 CM AGB/CDC-23122-XU1A, HEMODIALYSIS KIT: 2L 12 FR X 20 CM AGB/CDC-25122-1A, HEMODIALYSIS KIT: 2L 12 FR X 20 CM AGB/CDC-25122-X1A, AHDC AGB KIT: 2-L 12 FR X 20 CM/CDC-25122-XCN1A, AHDC AGB U-BEND KIT: 2-L 12 FR X 20 CM/CDC-25122-XCN1A-U, AHDC AGB KIT: 2-L 12 FR X 20 CM/CDC-25122-XN1A, HEMODIALYSIS KIT: 2L 12 FR X 20 CM AGB/CDC-25122-XU1A, HEMODIALYSIS KIT 2L: 14 FR X 20 CM AGB/CDC-25142-1A, HEMODIALYSIS KIT: 2L 14 FR X 20 CM AGB/CDC-25142-XC1A, AHDC AGB KIT: 2-L 14 FR X 20 CM/CDC-25142-XCN1A, AHDC AGB CURVED KIT: 2-L 14 FR X 20 CM/CDC-25142-XCN1A-C, AHDC AGB KIT: 2-L 14 FR X 20 CM/CDC-25142-XN1A, AHDC AGB CURVED KIT: 2-L 14 FR X 20 CM/CDC-25142-XN1A-C, HEMODIALYSIS KIT: 2L 12 FR X 25 CM AGB/CDC-26122-1A, AHDC AGB KIT: 2-L 12 FR X 25 CM/CDC-26122-XCN1A, AHDC AGB KIT: 2-L 12 FR X 25 CM/CDC-26122-XN1A, HEMODIALYSIS KIT 2L: 14 FR X 25 CM AGB/CDC-26142-1A, HEMODIALYSIS KIT: 2L 14 FR X 25 CM AGB/CDC-26142-X1A, AHDC AGB KIT: 2-L 14 FR X 25 CM/CDC-26142-XCN1A, AHDC AGB KIT: 2-L 14 FR X 25 CM/CDC-26142-XN1A, HEMODIALYSIS KIT: 2-LUMEN 12 FR X 20 CM/AK-25122-F, AHDC AGB KIT: 2-L 12 FR X 16 CM/CDC-22122-XCN1A, LBCVC AGB KIT: 3-L 12 FR X 16 CM/CDC-22123-XCN1A, LBCVC AGB KIT: 3-L 12 FR X 20 CM/CDC-25123-XCN1A, LBCVC AGB KIT:3-L 12 FR X 20 CM/CDC-25123-XN1A

La agencia publicó este retiro solo en inglés. El nombre del producto y los códigos del paquete se muestran tal como los escribió la agencia, para que coincidan exactamente con su paquete.

Revise su paquete

REF(Material)/UDI-DI/Lot(Batch): AK-15122-F/10801902197136, 10801902193275, 10801902168525/33F24E0348, 33F24J0280, 33F24K0119, 33F24L0601, 33F25A0477, 33F25C0908, 33F24A0850; ASK-12123-UOC1/10801902193275/33F25B0430, 33F25D0119; ASK-12123-UPM/10801902193275, 10801902222463/33F25E0535, 33F25G0188, 33F25G0189, 33F26C0333, 33F25D0262; ASK-15123-UOC1/10801902193275/33F25C0378, 33F25D0121; ASK-15123-UPM/10801902193275, 10801902159950/33F25E0718, 33F25H0447, 33F26A1389, 33F24E0649; CDC-12123-P1A/40801902111768, 10801902193275/33F24D0632, 33F24F0758, 33F24L0465, 33F25A0281, 33F25E0794; CDC-12123-XP1A/10801902193275, 30801902111785, 10801902213416/33F25J0723, 33F26A1382, 33F26B0433, 36F25A0007, 36F25B0009, 36F25E0018, 33F24D0299; CDC-12123-XPCN1A/10801902213416, 10801902193275/33F24D0950, 33F24K0417, 33F24L0466, 33F25A0216, 33F25B0667, 33F25D0436, 33F25D0666, 33F25E0693, 33F25G0239, 33F25J0247, 33F26A1556, 33F26B0801, 33F26C0767, 33F25B0668; CDC-12123-XPN1A/10801902193275, 10801902213409/33F25D0554, 33F25G0240, 33F25J0312, 33F26B0214,33F26C0199, 33F25A0181; CDC-15123-P1A/10801902193275/33F25D0437, 33F25H0547,33F25B0613; CDC-15123-XPCN1A/10801902213430, 10801902193275/33F24F0107, 33F24F0998, 33F24K0403, 33F24L0834, 33F25A0283, 33F25B0527, 33F25E0650, 33F25E0745, 33F25F0085, 33F25H0596, 33F25J0314, 33F26A1464, 33F26B0883, 33F25B0669; CDC-15123-XPN1A/0801902193275, 10801902213423, 10801902116793/33F25D0439, 33F25G0238, 33F25H0597, 33F25J0315, 33F26A1468, 33F26B0826, 33F24C0461; CDC-16123-XPCN1A/10801902193275/33F25B0671, 33F25D0440, 33F25E0651, 33F25G0236, 33F25H0529, 33F25J0316, 33F25G0237, AK-25122-CDC/10801902193275/33F24G0370, 33F24K0135, 33F25C0120, 33F25C0969, 33F25E0942, 33F24G0371; AK-25142-F/10801902193275, 70801902123205, 10801902167795/33F25C0863, 33F25E0087, 33F25H0038, 33F25J0846, 33F26B0531, 33F24D0089; ASK-22122-MM1/10801902193275/33F25E0546; ASK-22122-MM2/10801902193275, 10801902221923, 10801902144956/33F25G0172, 33F25L0031, 33F25M0310, 33F26A1652, 33F26B0151, 33F26B0450, 33F24F0854; ASK-22142-UOC1/10801902193275/33F25C0545, 33F25D0122; ASK-22142-UPM/10801902193275, 10801902222487, 10801902202281/33F25D0594, 33F25F0116, 33F25G0187, 33F26A1390, 33F26B0205, 33F24D0396; ASK-25142-UNC1/10801902193275/33F25C0655, 33F24K0561; ASK-25142-UOC1/10801902193275/33F25B0362, 33F25D0123; ASK-25142-UPM/10801902193275/33F25E0595, 33F25H0446, 33F25J0381, 33F26A1446, 33F25J0092; ASK-26142-UNC1/10801902193275/33F25B0427, 33F25K0058; AU-22122-F/10801902193275/33F24K0269, 33F25A0584, 33F25C0862, 33F25E0245, 33F25J0567; CDC-22122-1A/20801902123484, 10801902193275/, CDC-22122-XCN1A-U/33F24F0996, 33F24L0374, 33F25C0461; CDC-22122-XN1A/10801902193275/33F25C0239, 33F25E0875, 33F25J0419; CDC-22122-XN1A-U/10801902193275, 10801902213621/33F25A0309, 33F25E0876, 33F26B0217; CDC-22142-1A/50801902123515, 10801902193275/33F24D0928, 33F24F0103, 33F24F0896, 33F24L0111, 33F24L0955, 33F25B0500, 33F25D0447, 33F25E0654, 33F25G0278, 33F25H0577, 33F25J0319; CDC-22142-XC1A/10801902193275/33F25A0284, 33F25J0180; CDC-22142-XCN1A/10801902193275, 10801902213539/33F25A0285, 33F25B0529, 33F25D0449, 33F25E0655, 33F25E0818, 33F25F0214, 33F25G0233, 33F26A1477, 33F26B0898, 33R25J0455; CDC-22142-XCN1A-C/10801902193275/33F24L0473, 33F25B0615, 33F25F0213; CDC-22142-XN1A/10801902193275, 10801902213522/33F25C0240, 33F25E0578, 33F25F0215, 33F25H0528, 33F25J0321, 33F26A1383, 33F26B0215, 33F26C0016; CDC-22142-XN1A-C/10801902193275/33F25B0485, 33F25E0879; CDC-23122-XCN1A-U/10801902193275/33F24L0279, 33F25D0492; CDC-23122-XN1A-U/10801902193275/33F25E0579; CDC-23122-XU1A/10801902193275/33F25D0493, 33F25H0432, 33F25J0248; CDC-25122-1A/10801902193275/33F25B0619, 33F25E0881; CDC-25122-X1A/10801902193275/33F25E0882, 33F25J0422; CDC-25122-XCN1A/10801902213492, 10801902193275/33F24F0929, 33F25C0242, 33F25D0498, 33F25E0883, 33F25G0234, 33F26B0439, 33F26C0214; CDC-25122-XCN1A-U/10801902193275/33F24L1041, 33F25B0620, 33F25F0121, 33F25G0226; CDC-25122-XN1A/10801902193275, 10801902213485/33F25C0243, 33F25E0884, 33F25K0183, 33R25E0884; CDC-25122-XU1A/10801902193275/33F25B0621, 33F25E0886, 33F25H0568; CDC-25142-1A/50801902123584, 10801902193275/33F24F0837, 33F24K0457, 33F24L0113, 33F24L1031, 33F25B0622, 33F25D0503, 33F25E0889, 33F25F0106, 33F25J0127; CDC-25142-XC1A/10801902193275/33F25B0467; CDC-25142-XCN1A/10801902193275, 10801902213553/33F25A0312, 33F25E0657, 33F25E0821, 33F25G0231, 33F25H0527, 33F25J0428, 33F26B0452, 33F26C0133; CDC-25142-XCN1A-C/10801902193275/33F24L0377, 33F25D0238; CDC-25142-XN1A/10801902193275, 10801902213546/33F25B0624, 33F25D0504, 33F25E0823, 33F25G0232, 33F26B0437; CDC-25142-XN1A-C/10801902193275/33F24L0485, 33F25B0676; CDC-26122-1A/60801902123611, 10801902193275/33F24F0084, 33F24K0475, 33F25G0389; CDC-26122-XCN1A/10801902213515, 10801902193275/33F24F1028, 33F24L0117, 33F25D0308, 33F25D0557, 33F25F0219; CDC-26122-XN1A/10801902193275/33F25E0451; CDC-26142-1A/10801902193275/33F24L0488, 33F24L1025, 33F25C0368, 33F25E0891, 33F25H0406, 33F25J0116; CDC-26142-X1A/10801902193275/33F25D0558, 33F25E0892; CDC-26142-XCN1A/10801902193275, 10801902213577/33F24L0941, 33F25B0530, 33F25D0507, 33F25E0699, 33F25H0525, 33F26B0396, 33F26B0895, 33F26C0780; CDC-26142-XN1A/10801902193275, 10801902213560/33F24L0927, 33F25E0580, 33F25H0526, 33F26B0393; AK-25122-F/ CDC-22122-XCN1A/10801902193275/33F24G0371, 33F24L0742, 33F25C0822, 33F25E0084, 33F25J0483, 33F25L0400; CDC-22123-XCN1A/10801902213591, 10801902193275/33F24D0842,33F24F1018, 33F24L0110, 33F25C0205, 33F25D0613, 33F25E0878, 33F26B0455; CDC-25123-XCN1A/10801902213614, 10801902193275/33F24D0818, 33F24L0483, 33F25C0244, 33F25D0502, 33F25E0888, 33F25G0228, 33F25J0425, 33F26B0395, 33F26B0894; CDC-25123-XN1A/10801902193275/33F25C0332, 33F25G0229
Motivo
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
Distribuido
US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.
Empresa
ARROW INTERNATIONAL, LLC
Anunciado
El retiro comenzó
Leer el aviso oficial

Cada retiro — de alimentos, medicamentos o dispositivos médicos — se muestra palabra por palabra del aviso oficial de la FDA o del USDA, con un enlace directo. Aquí nada se resume, se reescribe ni se rellena. Este sitio no está afiliado a ninguna agencia del gobierno.

Local Recall