English
TemporalFDAEn curso

Puede enfermarle por un tiempo. No lo coma.

Philips Azurion 7M12, Model Numbers: 722078, 722223 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key

La agencia publicó este retiro solo en inglés. El nombre del producto y los códigos del paquete se muestran tal como los escribió la agencia, para que coincidan exactamente con su paquete.

Revise su paquete

UDI (01)00884838099241(21)85, (01)00884838099241(21)312, (01)00884838099241(21)332, (01)00884838099241(21)533, (01)00884838099241(21)479, (01)00884838099241(21)189, (01)00884838099241(21)47, (01)00884838085251(21)840, (01)00884838099241(21)352, (01)00884838099241(21)681, (01)00884838099241(21)391, (01)00884838099241(21)328, (01)00884838085251(21)870, (01)00884838099241(21)388, (01)00884838099241(21)581, (01)00884838099241(21)570, (01)00884838099241(21)77, (01)00884838099241(21)448, (01)00884838099241(21)571, (01)00884838099241(21)495, (01)00884838085251(21)679, (01)00884838099241(21)439, (01)00884838099241(21)628, (01)00884838099241(21)36, (01)00884838099241(21)580, (01)00884838099241(21)799, (01)00884838099241(21)577, (01)00884838085251(21)513, (01)00884838085251(21)833, (01)00884838085251(21)566, (01)00884838099241(21)323, (01)00884838099241(21)252, (01)00884838099241(21)626, (01)00884838099241(21)261, (01)00884838099241(21)125, (01)00884838099241(21)207, (01)00884838099241(21)329, (01)00884838099241(21)560, (01)00884838099241(21)561, (01)00884838099241(21)205, (01)00884838099241(21)379, (01)00884838099241(21)440, (01)00884838099241(21)204, (01)00884838099241(21)192, (01)00884838099241(21)197, (01)00884838085251(21)921, (01)00884838099241(21)177, (01)00884838085251(21)700, (01)00884838099241(21)37, (01)00884838099241(21)340, (01)00884838099241(21)357, (01)00884838099241(21)282, (01)00884838099241(21)214, (01)00884838085251(21)1007, (01)00884838099241(21)44, (01)00884838085251(21)862, (01)00884838099241(21)210, (01)00884838099241(21)336, (01)00884838099241(21)206, (01)00884838099241(21)343, (01)00884838085251(21)786, (01)00884838099241(21)365, (01)00884838099241(21)390, (01)00884838099241(21)186, (01)00884838099241(21)187, (01)00884838099241(21)631, (01)00884838099241(21)87, (01)00884838099241(21)373, (01)00884838099241(21)514, (01)00884838099241(21)69, (01)00884838099241(21)256, (01)00884838099241(21)679, (01)00884838099241(21)29, (01)00884838099241(21)67, (01)00884838099241(21)503, (01)00884838099241(21)432, (01)00884838099241(21)131, (01)00884838099241(21)704, (01)00884838099241(21)301, (01)00884838099241(21)456, (01)00884838085251(21)803, (01)00884838099241(21)659, (01)00884838099241(21)242, (01)00884838099241(21)746, (01)00884838099241(21)171, (01)00884838099241(21)457, (01)00884838099241(21)455, (01)00884838099241(21)221, (01)00884838099241(21)454, (01)00884838099241(21)318, (01)00884838099241(21)229, (01)00884838099241(21)230, (01)00884838099241(21)262, (01)00884838085251(21)890, (01)00884838085251(21)626, (01)00884838085251(21)697, (01)00884838099241(21)308, (01)00884838099241(21)124, (01)00884838099241(21)198, (01)00884838099241(21)433, (01)00884838085251(21)684, (01)00884838099241(21)566, (01)00884838099241(21)126, (01)00884838085251(21)829, (01)00884838099241(21)880, (01)00884838085251(21)143, (01)00884838099241(21)665, (01)00884838099241(21)188, (01)00884838085251(21)824, (01)00884838085251(21)649, (01)00884838099241(21)367, (01)00884838099241(21)52, (01)00884838099241(21)371, (01)00884838099241(21)127, (01)00884838099241(21)105, (01)00884838099241(21)572, (01)00884838099241(21)104, (01)00884838099241(21)334, (01)00884838099241(21)627, (01)00884838099241(21)458, (01)00884838099241(21)487, (01)00884838099241(21)463, (01)00884838099241(21)470, (01)00884838085251(21)902, (01)00884838099241(21)96, (01)00884838099241(21)277, (01)00884838099241(21)211, (01)00884838085251(21)317, (01)00884838085251(21)709, (01)00884838099241(21)97, (01)00884838099241(21)161, (01)00884838099241(21)386, (01)00884838099241(21)208, (01)00884838099241(21)361, (01)00884838099241(21)258, (01)00884838099241(21)597, (01)00884838085251(21)725, (01)00884838085251(21)728, (01)00884838099241(21)180, (01)00884838085251(21)441, (01)00884838099241(21)668, (01)00884838099241(21)377, (01)00884838099241(21)882, (01)00884838085251(21)906, (01)00884838099241(21)565, (01)00884838099241(21)223, (01)00884838099241(21)196, (01)00884838099241(21)130, (01)00884838099241(21)453, (01)00884838099241(21)452, (01)00884838099241(21)269, (01)00884838085251(21)788, (01)00884838099241(21)567, (01)00884838099241(21)449, (01)00884838085251(21)141, (01)00884838085251(21)279, (01)00884838085251(21)644, (01)00884838099241(21)435, (01)00884838099241(21)441, (01)00884838085251(21)844, (01)00884838085251(21)876, (01)00884838085251(21)748, (01)00884838099241(21)669, (01)00884838099241(21)351, (01)00884838099241(21)508, (01)00884838099241(21)750, (01)00884838099241(21)164.
Motivo
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Distribuido
US Nationwide distribution.
Empresa
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Anunciado
El retiro comenzó
Leer el aviso oficial

Cada retiro — de alimentos, medicamentos o dispositivos médicos — se muestra palabra por palabra del aviso oficial de la FDA o del USDA, con un enlace directo. Aquí nada se resume, se reescribe ni se rellena. Este sitio no está afiliado a ninguna agencia del gobierno.