TemporalFDAEn curso
Puede causar daños que por lo general son pasajeros.
Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Model Number: CI1000 UDI-DI codes: 9321502069847 Cochlear Nucleus CI1000 Series implants are prescription only, single use devices intended for long term implantation under the skin in the mastoid region on either side of the head. The cochlear implant system works by converting sound in the environment into electrical code. It transmits this code to the auditory nerve and on to the brain where it is interpreted as sound. The Cochlear Implant System consists of internal (implanted) and external components. The implanted component consists of a receiver/stimulator and an electrode array. The external components consist of a sound processor and their associated transmitting coils and cables and must be worn for the implant to work.
La agencia publicó este retiro solo en inglés. El nombre del producto y los códigos del paquete se muestran tal como los escribió la agencia, para que coincidan exactamente con su paquete.
Revise el dispositivo o su empaque
Model: CI1000 Product/Catalog: P1979576 UDI-DI code: 9321502069847 Serial Numbers: 1020210051972 1020210070461 1020210145046 1020210048286
- Motivo
- Due to a manufacturing deviation of incomplete electrode welds of cochlear implants and resulting in one or more open circuitsLa agencia escribió este motivo solo en inglés.La agencia escribió este motivo solo en inglés.
- Distribuido
- U.S.: NJ, NY, and PA. O.U.S.: Not provided
- Empresa
- Cochlear Limited
- Anunciado
- Esta es la fecha en que se anunció el retiro. La empresa pudo haberlo iniciado semanas antes.
- El retiro comenzó
Fuente: registro Z-0349-2027 de FDA · consultado el 3 de noviembre de 2026
Leer el aviso oficial