Puede causar daños que por lo general son pasajeros.

Techno-Path Multichem AE Unassayed Level 1 REF:AE621X

La agencia publicó este retiro solo en inglés. El nombre del producto y los códigos del paquete se muestran tal como los escribió la agencia, para que coincidan exactamente con su paquete.

Revise el dispositivo o su empaque

UDI: (01)05391523442800/Lot # 65705261
Motivo
A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.La agencia escribió este motivo solo en inglés.La agencia escribió este motivo solo en inglés.
Distribuido
US Nationwide distribution in the state of Massachusetts.
Empresa
TECHNO-PATH MANUFACTURING LTD.
Anunciado
Esta es la fecha en que se anunció el retiro. La empresa pudo haberlo iniciado semanas antes.
El retiro comenzó

Fuente: registro Z-0205-2027 de FDA · consultado el 22 de octubre de 2026

Leer el aviso oficial

Cada retiro — de alimentos, medicamentos o dispositivos médicos — se muestra palabra por palabra del aviso oficial de la FDA o del USDA, con un enlace directo. Aquí nada se resume, se reescribe ni se rellena. Este sitio no está afiliado a ninguna agencia del gobierno.